人工心脏瓣膜体外耐久性试验流场研究Flow Field Investigation of Prosthetic Heart Valves during In Vitro Durability Testing
吴坤能,郭剑,颜文涛
摘要(Abstract):
根据GB/T 12279系列标准搭建人工心脏瓣膜耐久性试验条件下的流场试验平台,运用数字粒子图像测速系统分析加速磨损测试(accelerated wear testing,AWT)、动态失效模式测试(dynamic failure mode testing,DFM)及实时磨损测试(real-time wear testing,RWT)的流场特征。研究结果显示,AWT试验中瓣膜流出道射流呈团状,随频率升高(15~20 Hz),峰值流速、VSS峰值和RSS峰值均降低;DFM试验结果显示,压力梯度增加(100、140、180 mmHg)与峰值流速及RSS峰值呈正相关,VSS峰值无显著规律;RWT试验所观察到的流场更接近生理状态,射流充分发展,与AWT和DFM试验的流场存在显著差异,单一试验均无法全面反映瓣膜在实际植入环境中所面临的复杂力学环境。该研究为论证AWT、DFM和RWT对瓣膜耐久性进行综合评估的必要性提供了相关血流动力学依据。
关键词(KeyWords): 加速磨损试验;动态失效模式试验;实时磨损试验;数字粒子图像测速;流场特征
基金项目(Foundation): 上海市药品监督管理局课题研究项目(ZD-2024-09)
作者(Author): 吴坤能,郭剑,颜文涛
参考文献(References):
- [1]O'Brien K D. Pathogenesis of calcific aortic valve disease:a disease process comes of age(and a good deal more)[J]. Arterioscler Thromb Vasc Biol,2006,26(8):1721-1728.
- [2]Nkomo V T,Gardin J M,Skelton T N,et al. Burden of valvular heart diseases:a population-based study[J]. Lancet,2006,368(9540):1005-1011.
- [3]中国医师协会心血管内科医师分会结构性心脏病学组,潘文志,宋光远,等.三尖瓣反流经导管治疗的中国专家共识[J].中国介入心脏病学杂志,2024,32(10):551-561.
- [4]潘文志,方雁行,李捷,等.经导管主动脉瓣置换术2023年度报告[J].中国胸心血管外科临床杂志,2024,31(4):498-503.
- [5]张长东,钟禹成,李庚,等. 2024年先天性心脏病介入治疗年度报告[J].中国胸心血管外科临床杂志,2025,32(7):909-918.
- [6]Arnold S V, Reynolds M R, Wang Kaijun, et al.Health status after transcatheter or surgical aortic valve replacement in patients with severe aortic stenosis at increased surgical risk:results from the corevalve US pivotal trial[J]. JACC Cardiovasc Interv,2015,8(9):1207-1217.
- [7]Popma J J, Deeb G M, Yakubov S J, et al.Transcatheter aortic-valve replacement with a self-expanding valve in low-risk patients[J]. N Engl J Med,2019,380(18):1706-1715.
- [8]Russo M, Taramasso M, Guidotti A, et al. The evolution of surgical valves[J]. Cardiovasc Med,2017,20(12):285-292.
- [9]Capodanno D, Petronio A S, Prendergast B, et al.Standardized definitions of structural deterioration and valve failure in assessing long-term durability of transcatheter and surgical aortic bioprosthetic valves:a consensus statement from the European Association of Percutaneous Cardiovascular Interventions(EAPCI)endorsed by the European Society of Cardiology(ESC)and the European Association for Cardio-Thoracic Surgery(EACTS)[J]. Eur J Cardiothorac Surg,2017,52(3):408-417.
- [10]Dvir D,Bourguignon T,Otto C M,et al. Standardized definition of structural valve degeneration for surgical and transcatheter bioprosthetic aortic valves[J]. Circulation,2018,137(4):388-399.
- [11]ISO 5840-1:2021 Cardiovascular implants-Cardiac valve prostheses—Part 1:general requirements[S].
- [12]ISO 5840-3:2021 Cardiovascular implants—Cardiac valve prostheses—Part 3:heart valve substitutes implanted by transcatheter techniques:Geneva, Switzerland:International Organization for Standardization[S].
- [13]GB/T 12279.1—2024心血管植入器械人工心脏瓣膜第1部分:通用要求[S].
- [14]GB/T 12279.2—2024心血管植入器械人工心脏瓣膜第2部分:外科植入式人工心脏瓣膜[S].
- [15]GB/T 12279.3—2025心血管植入器械人工心脏瓣膜第3部分:经导管植入式人工心脏瓣膜[S].
- [16]Clark R E, Swanson W M, Kardos J L, et al.Durability of prosthetic heart valves[J]. Ann Thorac Surg,1978,26(4):323-335.
- [17]Sathananthan J, Hensey M, Landes U, et al. Longterm durability of transcatheter heart valves:insights from bench testing to 25 years[J]. JACC Cardiovasc Interv,2019,13(2):235-249.
- [18]Sailahari V P, Steven D, Michael S S, et al.Transcatheter heart valve downstream fluid dynamics in an accelerated evaluation environment[J]. Ann Biomed Eng,2021,49(9):2170-2182.
- [19]Barannyk O, Karri S, Oshkai P. Experimental investigation of pulsatile flow through prosthetic heart valves[J]. ASME,2013,4:1-10.
- [20]Gunning P S,Saikrishnan N,McNamara L M,et al.An in vitro evaluation of the impact of eccentric deployment on transcatheter aortic valve hemodynamics[J]. Ann Biomed Eng, 2014, 42(6):1195-1206.
- [21]Hatoum H,Moore B L,Maureira P,et al. Aortic sinus flow stasis likely in valve-in-valve transcatheter aortic valve implantation[J]. J Thorac Cardiovasc Surg,2017,154(1):32-43.
- [22]Madukauwa-David I D,Sadri V,Midha P A,et al. An evaluation of the influence of coronary flow on transcatheter heart valve neo-sinus flow stasis[J]. Ann Biomed Eng,2020,48(1):169-180.
- [23]T/CSBME 088—2025数字粒子图像测速(DPIV)测试人工心脏瓣膜流场的方法[S].
- [24]颜文涛.经导管主动脉瓣置换术血流动力学体外实验研究[J].中国医疗器械杂志,2023,47(4):383-390.
- [25]Yan Wentao, Li Jianming, Zhang Bowen, et al.Patient-specific bicuspid aortic valve hemodynamics study based on computer simulation and in vitro experiment[J]. Acta Mechanica Sinica, 2024, 40:34022.
- [26]Yan Wentao, Li Jianming, Wang Wenshuo, et al. A fluid-structure interaction study of different bicuspid aortic valve phenotypes throughout the cardiac cycle[J]. Fronti Physiol,2021,12:716015.
- [27]巩雪纯,王封,郗恒东,等. PIV空间分辨率对湍流多点统计量测量的影响[J].实验流体力学,2024,38(4):44-57.
- [28]Catalano C,Pasta S,Potratz P,et al. Haemodynamic performance of transcatheter heart valve in bileaflet mechanical valve:an in-vitro study[J]. Interdiscip Cardiovasc Thorac Surg,2025,40(11):ivaf265.